中文版 | ENGLISHGuangzhou Lingjie Air Purification Equipment Manufacturing|GZLJ|air fiter|HEPA|Activated carbon filter |air shower|clean workbench|FFU|Passbox
4008-727374
18925063248
Online Service
Sampling car
【Name】:Sampling car
<< Former  Next >>

Sampling car
Product application
The product is suitable for use in pharmaceutical formulations (sterile except) sampling of raw materials. According to the requirements of GMP standard, sampling vehicle production of the company reached class clean degree, electric components and using GB1497 - 85 (electric safety standards); with GB--191 (packaging standards) regulations.
Working principle
Clean sampling car uses a vertical laminar flow (one-way flow) air flow organization, variable speed centrifugal fan to the suction box through the coarse filter pre filtered air into the plenum, then through the effective filter two filter. The wind blown clean air from the high efficiency filter, at a uniform velocity through the work area, the dust particles and microbial particles away, thereby forming a clean working environment. For the convenience of users, the detachable table, when the sampling weight of material is ≤ 10kg, can be carried out in the working face, the level indicator 180 can reach thousand grade purification standard. When the sampling material weight greater than or equal to 10kg, it must be on two pieces of work table removed, with material barrel with Dolly, the level indicator 220V.
Structural characteristics
The sample car box adopts whole steel structure form, box after degreasing, rust, phosphating process, surface spray treatment. The upper part of the box body is sending the bellows and static pressure tank, high efficiency filter with static pressure box. The fan system of adjustable air volume, light touch type switch voltage regulation size range, the average wind speed work area remains the ideal, so as to effectively prolong the service life of high efficiency air filter. 2537A short wave ultraviolet germicidal lamp using the sample vehicle radiation strong, active cells which can not only kill microorganisms, but also for the strong heat and other fungal spores of bacillus can play a role in killing. In addition, phages and viruses can also be destroyed quickly in the role of ultraviolet light.
Installation:
1, external connection solid sampling vehicle power supply, prevent connection is not real and burning fan.
2, sampling vehicle allows for short distances, but must guarantee the ground smooth, moving speed of not more than 6 m / min, in order to prevent the outside airflow interference with the work area cleanliness. Mobile power line should be of sufficient length, to prevent the power line appear to move in and cause accidents is broken!
Sampling vehicle basic parameters:
Clean grade million level ≥ 0.5 μ m (USA federal standard 209E) class @ ≥ 0.5 μ m (American federal standard 209
The average wind speed 0.25-0.6m/s 0.25-0.6m/s
Noise ≤ 58dB (A) ≤ 58dB (A)
Power supply AC, single phase 220V/50Hz AC, single phase 220V/50Hz
Maximum power 0.3kW 0.3kW
Normal working conditions of 5 ℃ ~ 40 ℃, relative humidity 80%: ≤ 5 ℃ ~ 40 ℃ relative humidity: ≤ 80%
Atmospheric pressure is 86 ~ 106KPa, 86 ~ 106KPa
Appearance size 700 x 1000 x 1800
Operational qualification:
Caster: flexible and reliable. Test method: observation, while hand test.
Illumination: the illumination shall be not less than 300Lx. Test method: by luminometer test, from the ground 60cm illumination measurement, measuring a total of three points, measured three times per point.
Purification efficiency
The number of dust particles: the number of particles larger than 5 μ m less than 20000 /m3, the number of particles larger than 0.5 μ m less than 3500000 /m3. Test method: adjust the speed to low, medium or high, tested using dust particle counter, 80cm-100cm from the ground point determination, each gear measuring three points, three measurements per poin
Settlement: settlement of bacteria bacteria ≤ 10, measuring a total of three points, three measurements per point.
Pressure: inside and outside pressure is greater than 10Pa. Test method: measuring the differential pressure micro differential pressure meter, measuring a total of three points, three measurements per point.
Technology and business consulting: 18925063248

Related products: car flow sampling sampling car clean sampling vehicle medical laboratory sampling sampling car car car car Shenzhen Guangzhou sampling sampling sampling vehicle manufacturers sampling car wholesale sampling vehicle manufacturing


From:Guangzhou Lingjie Air Purification Equipment Manufacturing Co.,Ltd. http://www.linjec.com   Date:2015-11-24   Hits:6511
  • Relate: